| Instrument identity | The tester’s model, serial number, and relevant options match the unit being offered. | Photograph of the identification label and the corresponding calibration certificate. | Confirm the serial number and configuration agree across the instrument, certificate, and quotation. |
| Calibration laboratory accreditation | The calibration was performed by a laboratory accredited to ISO/IEC 17025 for the relevant calibration work. | Accreditation certificate and the laboratory’s published scope of accreditation. | Check that the accreditation was valid on the calibration date and that the scope covers the applicable measurement types and ranges. |
| Measurement coverage | The certificate covers functions needed for the intended work, such as RF power, frequency, modulation, or audio measurements, where applicable. | Calibration points, ranges, methods, and results listed on the certificate or its attached data. | Compare the listed functions and ranges with the communication standards and test procedures you plan to use. |
| Metrological traceability | The laboratory documents an unbroken, documented calibration chain linking results to appropriate measurement references, with uncertainty stated at each relevant step. | Certificate traceability statement, measurement references, and uncertainty information. | Do not rely on a traceability statement alone if the certificate lacks results or uncertainty information needed to assess suitability. |
| Results and measurement uncertainty | The certificate reports actual calibration results and the associated measurement uncertainty, rather than only a pass statement, when those details are needed for the application. | As-measured values, units, uncertainty, and any stated coverage factor or confidence information. | Assess whether the uncertainty is appropriate for your required measurement tolerance and decision process. |
| As-found and as-left condition | The record distinguishes the instrument’s condition before any adjustment from its condition afterward, if adjustment or repair occurred. | As-found and as-left readings, adjustment notes, and any out-of-tolerance findings. | Ask whether out-of-tolerance results could affect prior measurements or indicate damage, drift, or an incomplete repair. |
| Dates and interval planning | The calibration date is clear, and any recommended or stated next-calibration date is identified. ISO/IEC 17025 does not prescribe one universal calibration interval for every instrument. | Calibration date, any due-date label, and available calibration or service history. | Set the interval according to use, stability history, environment, required accuracy, and your quality procedures. |
| Certificate completeness | The report identifies the laboratory, instrument, calibration date, procedure or method, results, and any relevant environmental conditions or limitations. | A complete, legible calibration certificate, including referenced attachments. | Check that pages and attachments are present and that the certificate has not been altered or separated from its identifying details. |
| Decision rule and conformity statement | If the certificate states conformity to a specification, the applied decision rule is clear where required and relevant to interpreting the result. | Stated specification limits, conformity outcome, and decision-rule information when applicable. | Confirm that the pass/fail basis matches your acceptance limits and accounts for uncertainty as required by your process. |
| Functional condition and service history | The tester powers up, completes self-tests, and has no unresolved faults, missing accessories, or unreported repairs that affect measurements. | Current self-test results, repair records, included accessories, and a demonstration or inspection report. | Verify essential functions before purchase and arrange recalibration after repair or transport if your risk assessment or procedure calls for it. |